On order

TOCADERO ONE, Automatic On-Line Total Organic Carbon Analyzer, HORIBA

TOCADERO ONE Horiba

The TOCADERO ONE is HORIBA's on-line total organic carbon analyzer for continuous TOC measurement in wastewater, industrial process water, and environmental water, manufactured by HORIBA Tocadero GmbH. Based on the 1200 °C combustion oxidation method without catalyst, it handles a wide concentration range from 0.1 ppm to 20,000 mg/L C and tolerates high-salinity samples up to 200 g/L TDS (20% NaCl). Fast cycle times (TC ≥1 min, NPOC ≥3 min), Condition Based Maintenance via integrated sensor monitoring, and a rich connectivity suite — including OPC UA, Profinet, Modbus, Bluetooth, WLAN, 5G, and ATEX/IECEx Zone 1/2 certification — make it suitable for demanding industrial and environmental monitoring applications.

Add to bag

The TOCADERO ONE is HORIBA's on-line TOC analyzer for continuous measurement of total organic carbon in wastewater, industrial effluent, process water, and environmental water, developed by HORIBA Tocadero GmbH. It is based on the 1200 °C high-temperature combustion oxidation method without catalyst: organic carbon compounds in the sample are completely oxidised in the combustion furnace, generating CO₂ that is measured by a detector to determine TOC concentration. The absence of a catalyst and the extreme combustion temperature ensure reliable oxidation even of refractory organic compounds, high-concentration samples (up to 20,000 mg/L C), and high-salinity matrices (up to 200 g/L TDS / 20% NaCl) with minimal furnace blockage. Fast cycle times — TC minimum 1 minute, NPOC minimum 3 minutes — support high-frequency process monitoring. Device status is continuously monitored by integrated sensors, enabling Condition Based Maintenance (CBM) for proactive servicing. Functions for reducing residue carryover effects and an automatic pure water cleaning feature maintain measurement integrity during flow path switching and concentration fluctuations. The highly scalable hardware and user-friendly 7" touch panel interface, combined with a comprehensive connectivity suite covering OPC UA, Profinet, Modbus, Bluetooth, WLAN, 5G, and ATEX/IECEx Zone 1/2 certification, make the TOCADERO ONE a versatile solution for demanding industrial environments.

Key Features TOCADERO ONE

  • 1200 °C combustion oxidation without catalyst — complete oxidation of refractory organics, high-concentration and high-salinity samples; reduced furnace blockage risk.
  • Wide measurement range: 0–1/100/1000/10,000 ppm (optional 0–100,000 ppm); LOD 0.1 ppm; handles TDS up to 200 g/L (20% NaCl) and particles up to 2000 µm (optional homogenizer and sample preparation).
  • Fast cycle times: TC ≥1 min; NPOC ≥3 min — suitable for high-frequency process monitoring.
  • Condition Based Maintenance (CBM) via continuous integrated sensor monitoring — proactive servicing and reduced unplanned downtime.
  • Automatic pure water cleaning and residue carryover reduction functions — stable measurements during concentration fluctuations and flow path switching.
  • 7" touch panel with multilingual interface (German, English, Japanese, Chinese, Korean).
  • Comprehensive connectivity: OPC UA, Ethernet, Profinet, Modbus, Bluetooth, WLAN, GSM, 5G; 4 programmable NAMUR relays.
  • ATEX/IECEx Zone 1/2, T3/T4 certification — suitable for hazardous area installation.
  • IP54/IP65 standard; optional NEMA4X for harsh environments.
  • Carrier gas: CO₂-free instrument air — no special gas supply required beyond clean dry air.

Applications

The TOCADERO ONE is suited to continuous on-line TOC monitoring in municipal and industrial wastewater treatment, factory effluent discharge monitoring, condensate contamination detection, airport and industrial drainage TOC management, environmental river and surface water monitoring, and any application requiring reliable high-concentration or high-salinity TOC measurement in demanding or potentially hazardous environments.

Installation and Safety Notes

Install indoors at 2–45 °C. Power: 110–230 V AC ±10 V, 50/60 Hz, approx. 600 VA. Carrier gas: CO₂-free instrument air. Particles in sample must be below 2000 µm (optional homogenizer available for samples with larger particles). For hazardous area installation, ATEX/IECEx Zone 1/2 T3/T4 certification applies. Consult HORIBA Tocadero for site-specific installation and integration requirements.

Technical Specifications

Parameter TOC (TC, NPOC)
Measurement method 1200 °C combustion oxidation (no catalyst)
Measurement ranges 0–1 ppm / 0–100 ppm / 0–1,000 ppm / 0–10,000 ppm (optional: 0–100,000 ppm)
Limit of detection 0.1 ppm
Repeatability ±2% end-of-range
Cycle time (TC) <1 min
Cycle time (NPOC/TOC) <3 min
TDS tolerance Up to 200 g/L (20% NaCl)
Particle size tolerance <2000 µm (optional homogenizer for larger particles)
Carrier gas CO₂-free instrument air
Display 7" touch panel; multilingual (DE/EN/JP/ZH/KO)
Communication protocols OPC UA; Ethernet; Profinet; Modbus
Wireless communication Bluetooth; WLAN; GSM; 5G
Relays 4 programmable — NAMUR standard
Power supply 110–230 V AC ±10 V, 50/60 Hz, approx. 600 VA
Operating temperature 2–45 °C (indoor)
Humidity <85% RH (non-condensing)
Protection rating IP54 / IP65; optional NEMA4X
Hazardous area certification ATEX / IECEx Zone 1/2, T3 and T4
Dimensions (H × W × D) 800 × 600 × 320 mm
Weight <55 kg
Manufacturer HORIBA Tocadero GmbH
Brand Horiba