HT-110, On-Line TOC Analyzer for Pharmaceutical Water, HORIBA

HT-110 Horiba

The HT-110 is HORIBA's on-line TOC analyzer designed for continuous measurement of total organic carbon in pharmaceutical manufacturing water — including Water for Injection (WFI) and Purified Water. Based on the continuous UV oxidation and conductivity method, it requires no reagents or carrier gases, delivers repeatability within ±2% FS, and complies with USP<643>, EP<2.2.44>, and JP<G8.4.5.2>. A deaeration mechanism suppresses false readings from air bubbles, and the extended UV lamp lifetime reduces maintenance to a single annual calibration.

Add to bag

The HT-110 is HORIBA's on-line TOC analyzer designed for continuous high-precision measurement of total organic carbon in pharmaceutical manufacturing water, including Water for Injection (WFI), Purified Water, and Highly Purified Water. It uses the continuous UV oxidation and conductivity method: organic carbon compounds in the sample are oxidised by UV irradiation, generating CO₂, which dissolves in the water and increases its conductivity — the TOC concentration is then calculated from the measured conductivity difference before and after oxidation. Because this method requires no reagents or carrier/purge gases, operating costs are significantly reduced and operational complexity is minimised. A built-in deaeration mechanism suppresses erroneous readings caused by the inclusion of air bubbles in the sample stream. The extended UV lamp lifetime reduces the maintenance schedule to a single annual calibration, minimising downtime in regulated pharmaceutical environments. An external contact input allows ON/OFF control for situations such as bypassing false alarms during SIP cleaning or nighttime shutdown of the purified water system. Installation options include pole mount, wall mount, and tabletop.

Key Features HT-110

  • Continuous UV oxidation and conductivity method — no reagents, no carrier gases; reduced operating expenses.
  • Measurement range 0–100/500/1000 ppb TOC; repeatability within ±2% FS; resolution 0.1 ppb (0–500 ppb range), 1 ppb (0–1000 ppb range).
  • Compliant with USP<643>, EP<2.2.44>, JP<G8.4.5.2> for WFI, Purified Water, and Highly Purified Water quality monitoring.
  • Built-in deaeration mechanism — suppresses false readings from air bubble inclusion in the sample.
  • Annual maintenance cycle — extended UV lamp lifetime reduces scheduled maintenance to once per year.
  • Contact input for external ON/OFF control — useful during SIP cleaning cycles or equipment shutdown without generating false alarms.
  • LCD display showing TOC, conductivity, water temperature, and alarm status simultaneously.
  • Single 4–20 mA DC output (or optional 0–1 V DC); two relay contact outputs for TOC/temperature/conductivity alarm and ON/OFF control.
  • Flexible installation: pole mount, wall mount, or tabletop.
  • IP43; 100–240 V AC; CE marked: EN61326-2-3, EN61010-1, RoHS EN50581; FCC Part 15.

Applications

The HT-110 is suited to continuous on-line TOC monitoring of Water for Injection (WFI), Purified Water, and Highly Purified Water in pharmaceutical manufacturing facilities — in compliance with USP<643>, EP<2.2.44>, and JP pharmacopoeial requirements. It is also applicable to TOC monitoring in other high-purity water applications where reagent-free low-maintenance continuous measurement is required.

Installation and Safety Notes

Install indoors (pole, wall, or tabletop). Power: 100–240 V AC ±10%, 50/60 Hz, 80 VA max.; built-in fuse 250 V AC 3.15 A (T). Sample conductivity must be below 10 μS/cm (below 2 μS/cm for the 0–100 ppb range); temperature 5–99 °C; pressure 0.03–0.50 MPa (fluctuation within ±0.05 MPa); flow rate 20–200 mL/min. Operating temperature: 5–40 °C; humidity below 85% RH.

Technical Specifications

Parameter TOC; Conductivity; Temperature (displayed)
Measurement method Continuous UV oxidation and conductivity method
Measurement range 0–100 ppb / 0–500 ppb / 0–1000 ppb TOC (selectable)
Repeatability Within ±2% FS
Display resolution 0.1 ppb (0–100 ppb and 0–500 ppb ranges); 1 ppb (0–1000 ppb range)
Sample conductivity <10 μS/cm (<2 μS/cm for 0–100 ppb range)
Sample temperature 5–99 °C
Sample pressure 0.03–0.50 MPa; fluctuation within ±0.05 MPa
Sample flow rate 20–200 mL/min
Transmission output 1 × 4–20 mA DC (isolated); optional 0–1 V DC; max. load 600 Ω
Contact output RLY1 Relay SPDT (1c); 125 V AC / 24 V DC, 3 A; upper/lower alarm, ON/OFF control; TOC, temperature, or conductivity
Contact output RLY2 Relay SPST (1a); 125 V AC / 24 V DC, 3 A; upper/lower alarm, ON/OFF control; TOC concentration
Contact input 1 × no-voltage 1a signal (open collector); insulated; external ON/OFF hold control
Display LCD with backlight; TOC, conductivity, temperature, alarm
Power supply 100–240 V AC ±10%, 50/60 Hz, 80 VA max.; fuse 250 V AC 3.15 A (T)
Operating temperature 5–40 °C
Humidity <85% RH
Protection rating IP43 (IEC60529 / JIS C0920); Pollution Level 2 (IEC61010-1)
Installation Indoor; pole mount, wall mount, or tabletop
Dimensions (W × H × D) 300 × 400 × 165 mm
Weight 13 kg
Compliance CE (EN61326-2-3, EN61010-1, RoHS EN50581); FCC Part 15; Pollution Level 2 IEC61010-1
Pharmacopoeia compliance USP<643>; EP<2.2.44>; JP<G8.4.5.2>
Manufacturer HORIBA Advanced Techno Co., Ltd.
Brand Horiba