The HE-960CW is a 96 × 96 mm DIN panel-mount 2-channel conductivity meter from HORIBA's 48/96 Series, specifically designed for pharmaceutical water quality management and SIP/CIP monitoring. It accepts two independent, isolated conductivity sensor inputs and simultaneously displays conductivity, TDS, temperature, desalination rate, and conductivity difference between channels. Water quality assessment is compliant with USP<645>, making it ideal for Water for Injection (WFI) and Purified Water management. Four isolated 4–20 mA outputs, four relay contact outputs with USP assessment alarm, and RS-485 communication complete the feature set.
HE-960CW, Panel-Mount Type 2-Channel Conductivity Meter, Sanitary and Pharmaceutical, HORIBA
The HE-960CW is a 2-channel panel-mount conductivity meter from HORIBA's 48/96 Series, purpose-built for pharmaceutical water quality management and sanitary process monitoring. It accepts two independent, electrically isolated conductivity sensor inputs and simultaneously monitors conductivity, TDS, temperature, desalination rate (0–100%), and conductivity difference between channels — making it ideal for reverse osmosis system monitoring, WFI loop management, and SIP/CIP phase verification. Water quality assessment is fully compliant with USP<645>, with a dedicated USP assessment alarm contact output. Four isolated 4–20 mA outputs and RS-485 support integration into pharmaceutical process automation systems. Compatible sensors include standard ESD/ESL/ESH/FS Series as well as sanitary insert and flow-through types compliant with USP and EP pharmacopoeias.
Key Features HE-960CW
- 2 independent isolated sensor input channels — simultaneous measurement at two process points with desalination rate and conductivity difference display.
- USP<645> compliant water quality assessment with dedicated alarm contact output — ideal for WFI, Purified Water, and pharmaceutical process monitoring.
- Compatible with sanitary insert sensors (ESD-01C-SN1.5, ESH-01C-SN1.5) and sanitary flow-through sensors (FS-01C-SL, FS-01FC-SL) in addition to standard ESD/ESL/ESH/FS Series.
- Four isolated 4–20 mA DC outputs (max. 900 Ω; 4–20 or 0–20 mA selectable): freely assignable to conductivity, TDS, temperature, desalination rate, or conductivity difference.
- Four relay contact outputs (R1–R3: SPST 1a; R4: SPDT 1c; 240 V AC / 30 V DC, 3 A): selectable as upper/lower limit alarm, ON/OFF control, USP assessment, anomaly alarm, or maintenance.
- RS-485 communication for SCADA/DCS integration.
- Temperature compensation: NaCl characteristic, arbitrary coefficient (±5%/°C), or none; auto deionised water compensation; range 0–100 °C.
- Panel face IP65 (IEC60529 / JIS C0920); ABS case; 96 × 96 mm DIN; Class II device IEC61010-1, Pollution Level 2.
- CE marked: EN61326-1, EN61010-1, RoHS EN50581; FCC Part 15.
Applications
The HE-960CW is suited to pharmaceutical water quality management (WFI, PW — USP<645> compliance), SIP temperature and conductivity monitoring, CIP rinse verification, reverse osmosis desalination rate monitoring, and any application requiring simultaneous dual-point conductivity measurement with pharmacopoeia-grade assessment in a panel-integrated configuration.
Installation and Safety Notes
Install into a standard 96 × 96 mm panel cutout. Power: 100–240 V AC, 50/60 Hz, 25 VA max. Panel face IP65; rear case IP20; terminals IP00 — protect the rear from moisture. Operating temperature −5 to 55 °C; storage −25 to 65 °C; humidity 20–85% RH non-condensing.
Technical Specifications
| Parameter | Conductivity, TDS, Temperature, Desalination Rate, Conductivity Difference |
| Measurement method | 2-electrode conductivity |
| Sensor inputs | 2 channels, electrically isolated from each other |
| Compatible sensors | ESD/ESL/ESH/FS Series; sanitary insert (ESD-01C-SN1.5, ESH-01C-SN1.5); sanitary flow-through (FS-01C-SL, FS-01FC-SL); cell const. 0.01, 0.1, 1.0 /cm |
| Conductivity range (cell const. 0.01 /cm) | 0–2.000 / 0–20.00 μS/cm |
| Conductivity range (cell const. 0.1 /cm) | 0–2.000 / 0–20.00 / 0–200.0 / 0–2000 μS/cm (FS-01 only for top range) |
| Conductivity range (cell const. 1.0 /cm) | 0–200.0 / 0–2000 μS/cm |
| Desalination rate | 0–100% |
| Temperature range | 0–100 °C (display −30 to 160 °C) |
| Repeatability / Linearity (conductivity) | Within ±0.5% FS; ±2% FS for FS-01 at 2000 μS/cm range |
| Repeatability / Linearity (TDS) | Within ±1.5% FS |
| Transmission output | 4 × 4–20 mA DC (or 0–20 mA), isolated; max. 900 Ω; free range setting |
| Contact output | R1–R3: SPST (1a), 240 V AC / 30 V DC, 3 A; R4: SPDT (1c), 240 V AC / 30 V DC, 3 A; USP assessment, upper/lower alarm, ON/OFF control, anomaly alarm, maintenance |
| Communication | RS-485 |
| USP compliance | USP<645> water quality assessment |
| Temperature compensation | NaCl / arbitrary coefficient ±5%/°C / none; auto DI water compensation; Pt 1 kΩ; range 0–100 °C |
| Power supply | 100–240 V AC, 50/60 Hz, 25 VA max. |
| Operating temperature | −5 °C to 55 °C |
| Storage temperature | −25 °C to 65 °C |
| Humidity | 20–85% RH (non-condensing) |
| Protection rating | Panel face IP65; rear case IP20; terminals IP00 (IEC60529 / JIS C0920) |
| Installation type | Indoor panel-mount (48/96 Series, DIN 96×96 mm) |
| Housing material | ABS (panel case); PBT (terminals) |
| Dimensions (W × H × D) | 96 × 96 × 115 mm (case depth approx. 105 mm) |
| Weight | Approx. 550 g |
| Compliance | CE (EN61326-1, EN61010-1, RoHS EN50581); FCC Part 15; Class II device IEC61010-1; Pollution Level 2 |
| Manufacturer | HORIBA Advanced Techno Co., Ltd. / Finesse Solutions, LLC |
| Brand | Horiba |