The OXYTRANS TR is an optical inline oxygen sensor from Centec for continuous measurement of dissolved O₂ in liquids and O₂ in gas phases, based on luminescence quenching technology. Unlike conventional electrochemical sensors, it requires no electrolyte solution and does not consume oxygen during measurement — delivering high long-term stability with minimal maintenance. Available in two measurement ranges for both liquid and gas phase applications, with automatic Pt100 temperature compensation, full CIP compatibility, and integration capability into Centec deaeration and water treatment systems including OXYSTEAM, DeGaS-Cold, DeGaS-Hot, and DGS.
The OXYTRANS TR is an optical inline O₂ sensor from Centec for continuous measurement of dissolved oxygen in liquids and oxygen in gas phases, using the luminescence quenching principle. An LED light source illuminates a thin indicator layer through an optical window; the layer emits red light that is detected inside the sensor. When oxygen molecules are present, they absorb excitation energy from the indicator layer, reducing the emitted light signal in proportion to O₂ concentration. Because only oxygen absorbs this energy, other gases present in the medium have no influence on the measurement result. Unlike electrochemical sensors, OXYTRANS TR requires no electrolyte replacement and does not consume oxygen during measurement — making it significantly lower-maintenance and more stable over extended operating periods. Two range variants are available for both liquid and gas phase measurements, covering ultra-low ppb-level dissolved oxygen in deaerated water through to high-percentage gas phase O₂. OXYTRANS TR can be deployed as a standalone inline sensor or integrated into Centec process systems including OXYSTEAM, DeGaS-Cold, DeGaS-Hot, and DGS.
Key Features OXYTRANS TR
- Optical luminescence quenching principle — no electrolyte, no O₂ consumption; high long-term measurement stability.
- Dual application: dissolved O₂ in liquids (deaerated water, wort, beer, CO₂) and O₂ in gas phase — two range variants per phase.
- Liquid phase: variant I 1 ppb–2 ppm (±1 ppb); variant II 30 ppb–35 ppm (±30 ppb).
- Gas phase: variant I 0–4.2% O₂ (±0.002%); variant II 0–50% O₂ (±0.03%).
- Response time t₉₀ ≤10 s — fast inline process feedback.
- Other gases present in the medium do not affect the measurement result.
- Automatic temperature compensation via integrated Pt100 sensor.
- Aseptic design; fully CIP-compatible; wetted materials 1.4404 / AISI 316L, silicone FDA, PTFE FDA.
- Varivent inline connection; other connections on request. IP65; 24 V DC.
- Outputs: 2 × 4–20 mA analogue; 2 × digital 24 V DC; Profibus DP optional.
- Optional high-temperature version: operating temperature up to +98 °C.
Applications
The OXYTRANS TR is suited to inline O₂ monitoring in deaerated water (water treatment systems), wort and beer production (dissolved O₂ control for flavour stability and shelf life), CO₂ purity monitoring, and boiler steam/gas phase O₂ measurement as part of the OXYSTEAM deaeration control system. It is also used in deaeration systems including DeGaS-Cold, DeGaS-Hot, and DGS.
Installation and Safety Notes
Install using Varivent inline connections or alternative connections as specified. Power: 24 V DC. Fully CIP-compatible. Working pressure max. 12 bar; standard operating temperature −5 to +55 °C; high-temperature option up to +98 °C. No electrolyte replacement is required. Consult the technical datasheet for full wiring and installation specifications.
Technical Specifications
| Measured parameter | O₂ (dissolved in liquid and in gas phase) |
| Measurement method | Optical luminescence quenching |
| Applications | Deaerated water; wort; beer; CO₂; steam/gas phase (OXYSTEAM) |
| Liquid phase range (variant I) | 1 ppb–2 ppm O₂; accuracy ±1 ppb |
| Liquid phase range (variant II) | 30 ppb–35 ppm O₂; accuracy ±30 ppb |
| Gas phase range (variant I) | 0–4.2% O₂; accuracy ±0.002% O₂ |
| Gas phase range (variant II) | 0–50% O₂; accuracy ±0.03% O₂ |
| Response time t₉₀ | ≤10 s |
| Temperature sensor | Pt100 (automatic temperature compensation) |
| Working pressure | Max. 12 bar |
| Operating temperature | −5 °C to +55 °C (option: up to +98 °C) |
| Wetted materials | 1.4404 / AISI 316L; silicone FDA; PTFE FDA |
| Process connection | Varivent inline; others on request |
| Analogue outputs | 2 × 4–20 mA |
| Digital outputs | 2 × 24 V DC |
| Fieldbus | Profibus DP (optional) |
| Protection rating | IP65 |
| Power supply | 24 V DC |
| Electrolyte replacement | Not required |
| O₂ consumption during measurement | None |
| CIP compatibility | Full CIP compatibility |
| Manufacturer | Centec GmbH |
| Brand | Centec |