For manganese evaluation in water samples.
Manganese Color Disc Test Kit measures manganese in the range from 0 to 0.70 mg/L Mn in water. Color Disc/PAN method is used to perform manganese determination. Model MN-PAN may be used for evaluation of 50 samples. The smallest increment is 0.05 mg/L Mn. Manganese Color Disc Test Set is required for manganese monitoring in the beverage, paper and textile industries, dye production and food processing.
Manganese Color Disc Test Set includes:
- Alkaline Cyanide Reagent (p/n 2122332)
- Ascorbic Acid Powder Pillows (p/n 1457799)
- PAN Indicator Solution, 0.1% (p/n 2122432)
- Square Bottle, with 25-mL mark (p/n 1704200)
- PAN Manganese Color disc, 0–0.7 mg/L (p/n 9266200)
- Color Comparator Box (p/n 173200)
- 177-mL Demineralizer Bottle (p/n 1429900)
- 18 mm Glass Viewing Tubes (p/n 173006)
Procedure for Mn measurement
Before starting analysis:
- It is required to clean all labware with 1:1 nitric acid solution and rinse fully with deionized water.
- It is recommended to digest the sample with heat and acid to determine total manganese.
- If the sample is very acidic or alkaline, adjust the sample pH to 4–5 with sodium hydroxide or sulfuric acid solution. Keep the pH below 5 to prevent the precipitation of manganese.
- Check the instructions for possible interfering substances and their levels.
Testing procedure
- Add deionized water or pretreated sample into the first bottle to the 25‑mL mark.
- Fill the second bottle to the 25‑mL mark with sample.
- Pour the content of Ascorbic Acid Reagent Powder Pillow to each bottle and swirl to mix.
- Pipet 1 mL of Alkaline-Cyanide Reagent and stir the content of each bottle.
- Add 1 mL of PAN Indicator Solution to each bottle and swirl to mix., an orange colour will develop.
- After 2 minutes (or 10 minutes if the iron concentration in the sample is more than 5 mg/L), fill the first tube to the 5 mL line with water from the first bottle and put into the left opening of the color comparator box.
- Fill the second tube to the first line with the prepared sample. Insert the second tube into the color comparator box.
- Set the color comparator box in front of the light, rotate the color disc to match the color and read the testing results in mg/L in the scale window.
Brand | Hach |
Parameter | Manganese |
Measuring range | 0 - 0.70 mg/L Mn |
Smallest Increments Steps | 0.05 |
Method | Color Disc/PAN method |
Model | MN-PAN |
Number of tests | 50 tests |