On order

M1100‑S00H Orbisphere Luminescent Dissolved Oxygen Sensor 0–40 ppm, 28 mm Orbisphere Fitting

M1100-S00H Orbisphere

The Orbisphere M1100‑S00H is a luminescent dissolved oxygen sensor designed for inline measurement of oxygen concentrations up to 40 ppm. It features a 28 mm fitting and delivers precise, stable readings for beverage, water, and industrial process applications.

Add to bag

Description

The Orbisphere M1100 Luminescent Dissolved Oxygen Sensor (Part Number M1100‑S00H) provides accurate inline DO measurement up to 40 ppm. Using luminescent optical technology without membranes or electrolytes, it ensures stable, low-maintenance operation. The 28 mm fitting is compatible with standard Orbisphere analyzers and suitable for beverage, water, and industrial process monitoring.

Advantages M1100‑S00H

  • Wide measurement range 0–40 ppm
  • Luminescent technology with minimal drift
  • No membrane or electrolyte, reducing maintenance
  • Fast response time and reliable long-term performance
  • 28 mm fitting for standard Orbisphere installation
  • Robust design resistant to CIP cleaning

Usage Instructions

Install the M1100‑S00H sensor in a compatible Orbisphere analyzer or controller. Perform calibration using oxygen-free gas (e.g., high-purity nitrogen). Follow standard CIP procedures for cleaning and refer to Hach documentation for storage and maintenance.

Technical Specifications

Parameter Specification
Measuring Range 0–40 ppm DO
Accuracy ± 0.02 ppm or ± 3 % of reading
Lower Detection Limit 0.015 ppm
Repeatability ± 0.015 ppm or ± 2 %
Response Time (90 %) < 10 s (gas), < 30 s (liquid)
Fitting 28 mm Orbisphere fitting
Operating Temperature ‑5 °C to 50 °C (operating), resistant ‑5 °C to 100 °C
Flow Rate 50–300 mL/min
Sample Pressure 1–20 bar absolute
Material (Wetted) 316 L Stainless Steel, Silicone
Certifications 2004/108/EC, EN 61326‑1
Weight 0.74 kg
Included with Sensor Sensor only; controller, cable, or flow chamber ordered separately
Brand Orbisphere